DC/DC Medical Converter
•
•
Compact DIP-16-package
•
Certification according to IEC/EN/ES
60601-1 3rd edition for 2xMOPP and
operation up to 5000 m altitude
TIM 2 Series, 2 Watt
I/O isolation 5000 VAC rated for 250
VAC working voltage
•
•
Low leakage current < 2 µA
•
5-year product warranty
Extended operating temperature range
–40°C to 95°C.
ES 60601-1 IEC 60601-1
UL 62368-1 IEC 62368-1
The TIM 2 series is a range of 2 Watt DC/DC converters in compact DIP-16
package with reinforced isolation of 5000 VAC for medical applications. With a
low leakage current of less than 2 µA the converters are predestined to insulate
electrical equipment from the applied parts to patient (BF classification). The
models are approved to IEC/EN/ES 60601-1 3rd ed. for 2xMOPP up to an altitude of 5000 m and come along with an ISO 14971 risk management file.
Models
Order Code
Input Voltage
Range
TIM 2-0910
TIM 2-0911
TIM 2-0919
TIM 2-0912
TIM 2-0913
TIM 2-0915
TIM 2-0922
TIM 2-0923
TIM 2-1210
TIM 2-1211
TIM 2-1219
TIM 2-1212
TIM 2-1213
TIM 2-1215
TIM 2-1222
TIM 2-1223
TIM 2-2410
TIM 2-2411
TIM 2-2419
TIM 2-2412
TIM 2-2413
TIM 2-2415
TIM 2-2422
TIM 2-2423
TIM 2-4810
TIM 2-4811
TIM 2-4819
TIM 2-4812
TIM 2-4813
TIM 2-4815
TIM 2-4822
TIM 2-4823
www.tracopower.com
4.5 - 12 VDC
(9 VDC nom.)
9 - 18 VDC
(12 VDC nom.)
18 - 36 VDC
(24 VDC nom.)
36 - 75 VDC
(48 VDC nom.)
Output 1
Output 2
Vnom
Imax
3.3 VDC
5 VDC
9 VDC
12 VDC
15 VDC
24 VDC
+12 VDC
+15 VDC
3.3 VDC
5 VDC
9 VDC
12 VDC
15 VDC
24 VDC
+12 VDC
+15 VDC
3.3 VDC
5 VDC
9 VDC
12 VDC
15 VDC
24 VDC
+12 VDC
+15 VDC
3.3 VDC
5 VDC
9 VDC
12 VDC
15 VDC
24 VDC
+12 VDC
+15 VDC
600 mA
400 mA
222 mA
167 mA
134 mA
83 mA
83 mA
67 mA
600 mA
400 mA
222 mA
167 mA
134 mA
83 mA
83 mA
67 mA
600 mA
400 mA
222 mA
167 mA
134 mA
83 mA
83 mA
67 mA
600 mA
400 mA
222 mA
167 mA
134 mA
83 mA
83 mA
67 mA
November 15, 2023
Vnom
Efficiency
Imax
–12 VDC
–15 VDC
83 mA
67 mA
–12 VDC
–15 VDC
83 mA
67 mA
–12 VDC
–15 VDC
83 mA
67 mA
–12 VDC
–15 VDC
83 mA
67 mA
typ.
75 %
78 %
78 %
82 %
82 %
82 %
82 %
80 %
76 %
78 %
79 %
82 %
82 %
81 %
81 %
81 %
76 %
79 %
80 %
81 %
81 %
81 %
81 %
81 %
76 %
78 %
79 %
80 %
82 %
81 %
81 %
81 %
Page 1 / 5
TIM 2 Series, 2 Watt
Input Specifications
Input Current
- At no load
9 Vin models:
12 Vin models:
24 Vin models:
48 Vin models:
9 Vin models:
12 Vin models:
24 Vin models:
48 Vin models:
9 Vin models:
12 Vin models:
24 Vin models:
48 Vin models:
9 Vin models:
12 Vin models:
24 Vin models:
48 Vin models:
Surge Voltage
Under Voltage Lockout
Recommended Input Fuse
Input Filter
80 mA typ.
40 mA typ.
25 mA typ.
12 mA typ.
15 VDC max. (1 s max.)
25 VDC max. (1 s max.)
50 VDC max. (1 s max.)
100 VDC max. (1 s max.)
2 VDC min. / 3 VDC typ. / 4 VDC max.
6 VDC min. / 7 VDC typ. / 8 VDC max.
13 VDC min. / 15 VDC typ. / 17 VDC max.
29 VDC min. / 32 VDC typ. / 35 VDC max.
1'000 mA (slow blow)
500 mA (slow blow)
315 mA (slow blow)
160 mA (slow blow)
(The need of an external fuse has to be assessed
in the final application.)
Internal Capacitor
Output Specifications
Voltage Set Accuracy
Regulation
- Input Variation (Vmin - Vmax)
- Load Variation (10 - 90%)
Ripple and Noise
Capacitive Load
- Cross Regulation
(25% / 100% asym. load)
- 20 MHz Bandwidth
- single output
- dual output
±1% max.
0.2% max.
0.2% max.
0.5% max.
0.8% max. (Output 1)
0.8% max. (Output 2)
dual output models: 5% max.
single output models:
dual output models:
single output models:
dual output models:
3.3 Vout models:
5 Vout models:
9 Vout models:
12 Vout models:
15 Vout models:
24 Vout models:
12 / -12 Vout models:
15 / -15 Vout models:
Minimum Load
Temperature Coefficient
Start-up Time
Short Circuit Protection
Overload Protection
Overvoltage Protection
Transient Response
- Response Time
50 mVp-p typ.
1'000 µF max.
1'000 µF max.
430 µF max.
220 µF max.
170 µF max.
100 µF max.
170 / 170 µF max.
100 / 100 µF max.
Not required
±0.02 %/K max.
10 ms typ. / 20 ms max.
Continuous, Automatic recovery
Foldback Mode
104 - 197% of Vout nom.
(depending on model)
4 - 6.5 VDC (3.3 Vout models)
6 - 8 VDC (5 Vout models)
10 - 14 VDC (9 Vout models)
13 - 19 VDC (12 Vout models)
16 - 22 VDC (15 Vout models)
25 - 35 VDC (24 Vout models)
500 µs typ. (25% Load Step)
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
November 15, 2023
Page 2 / 5
TIM 2 Series, 2 Watt
Safety Specifications
Standards
EN 62368-1
IEC 62368-1
UL 62368-1
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
www.tracopower.com/overview/tim2
PD 2
- IT / Multimedia Equipment
- Medical Equipment
- Certification Documents
Pollution Degree
EMC Specifications
EMI Emissions
- Conducted Emissions
- Radiated Emissions
External filter proposal:
EMS Immunity
- Electrostatic Discharge
Air:
Contact:
- RF Electromagnetic Field
- EFT (Burst) / Surge
Ext. input component:
- Conducted RF Disturbances
- PF Magnetic Field
Continuous:
1 s:
EN 60601-1-2 edition 4 (Medical Devices)
EN 55011 class B (with external filter)
EN 55032 class B (with external filter)
FCC Part 18 class B (with external filter)
EN 55011 class B (with external filter)
EN 55032 class B (with external filter)
FCC Part 18 class B (with external filter)
www.tracopower.com/overview/tim2
EN 60601-1-2 edition 4 (Medical Devices)
EN 61000-4-2, ±15 kV, perf. criteria A
EN 61000-4-2, ±8 kV, perf. criteria A
EN 61000-4-3, 10 V/m, perf. criteria A
EN 61000-4-4, ±2 kV, perf. criteria A
EN 61000-4-5, ±1 kV, perf. criteria A
9 Vin models: KY 1000 µF || TVS SMDJ18A
12 Vin models: KY 470 µF
24 Vin models: KY 470 µF
48 Vin models: KY 220 µF
EN 61000-4-6, 10 Vrms, perf. criteria A
EN 61000-4-8, 100 A/m, perf. criteria A
EN 61000-4-8, 1000 A/m, perf. criteria A
General Specifications
Relative Humidity
Temperature Ranges
Power Derating
Cooling System
Remote Control
- Operating Temperature
- Case Temperature
- Storage Temperature
- High Temperature
- Current Controlled Remote
(passive = on)
- Off Idle Input Current
Altitude During Operation
Regulator Topology
Switching Frequency
Insulation System
Working Voltage (rated)
Isolation Test Voltage
Creepage
Clearance
Isolation Resistance
- Input to Output, 60 s
- Input to Output
- Input to Output
- Input to Output, 500 VDC
95% max. (non condensing)
–40°C to +95°C
+105°C max.
–55°C to +125°C
6.67 %/K above 90°C
See application note: www.tracopower.com/overview/tim2
Natural convection (20 LFM)
On: open circuit
Off: 2 to 4 mA current (internal 1 kΩ resistor)
Refers to 'Remote' and '-Vin' Pin
External circuit proposal: www.tracopower.com/info/current-remote.pdf
2.5 mA typ.
5'000 m max.
Flyback Converter, RCC Converter
100 kHz min. (RCC)
Reinforced Insulation
250 VAC
5'000 VAC
8 mm min.
8 mm min.
10'000 MΩ min.
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
November 15, 2023
Page 3 / 5
TIM 2 Series, 2 Watt
Isolation Capacitance
- Input to Output, 100 kHz, 1 V
Leakage Current
Reliability
Washing Process
- Touch Current
- Calculated MTBF
Environment
- Vibration
- Mechanical Shock
- Thermal Shock
Housing Material
Base Material
Potting Material
Pin Material
Pin Foundation Plating
Pin Surface Plating
Housing Type
Mounting Type
Connection Type
Footprint Type
Soldering Profile
Weight
Environmental Compliance - REACH Declaration
- RoHS Declaration
- SCIP Reference Number
16 pF typ.
20 pF max.
2 µA max. (at 240 VAC / 60 Hz)
6'809'000 h (MIL-HDBK-217F, ground benign)
According to Cleaning Guideline
www.tracopower.com/info/cleaning.pdf
MIL-STD-810F
MIL-STD-810F
MIL-STD-810F
Non-conductive Plastic (UL 94 V-0 rated)
Non-conductive Plastic (UL 94 V-0 rated)
Silicone (UL 94 V-0 rated)
Copper
Nickel (1 - 3 µm)
Tin (7 - 12 µm), matte
Plastic Case
PCB Mount
THD (Through-Hole Device)
DIP16
Lead-Free Wave Soldering
260°C / 10 s max.
7g
www.tracopower.com/info/reach-declaration.pdf
REACH SVHC list compliant
REACH Annex XVII compliant
www.tracopower.com/info/rohs-declaration.pdf
Exemptions: 7a, 7c-I
(RoHS exemptions refer to the component
concentration only, not to the overall
concentration in the product (O5A rule).)
b0b10c00-0993-46b5-bf4c-c3885a5d658c
Supporting Documents
Overview Link (for additional Documents)
www.tracopower.com/overview/tim2
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
November 15, 2023
Page 4 / 5
TIM 2 Series, 2 Watt
3.2
14.4 (0.57)
10.16
(0.13)
Outline Dimensions
Pinout
(0.400)
9
2.1
(0.08)
Dual Output
–Vin (GND)
2
Remote
Remote
2
7
NC
NC
8
NC
Common
9
+Vout
+Vout
10
–Vout
–Vout
+Vin (Vcc)
+Vin (Vcc)
2.54
–Vin (GND)
(0.100)
1
BOTTOM VIEW
10
Single Output
1
15.24 (0.600)
24.3 (0.96)
16
Pin
16
7
8
NC: Not connected
Ø 0.60
(0.0024)
Dimensions in mm (inch)
Tolerances ±0.5 (±0.02)
Pin Ø 0.6 ±0.1 (0.024 ±0.004)
Pin pitch tolerances ±0.25 (±0.01)
3.8
(0.15)
10.2
(0.40)
© Copyright 2023 Traco Electronic AG
Specifications can be changed without notice.
www.tracopower.com
Rev. November 15, 2023
Page 5 / 5